AOI US Closed

Re: SPI question from locked ticket:1804851

CustomerAllen Hovey
CompanyALTERNATIVE MFG INC - WINTHROP
Priority1
TypeN/A
UpdatedAug 21, 2026 at 2:26 PM
Thread1 messages

Current case

Customer issue

Saqib Toor : Hello Temeka, Please run the 3D height verification jig to rule out any issues with the machine itself. If the 3D height verification is within 3 microns, please manually adjust the lighting settings for both the 3D inspection and solder paste inspection. The 3D lighting affects the height measurement, while the solder paste lighting affects the area and volume calculations. If the height verification passes, I would recommend reviewing the printer settings next

Conversation

Engineer and customer thread

1 messages
  1. Saqib Toor Private note | Aug 19, 2026 at 1:46 PM
    Private
    This ticket is closed and merged into ticket 1840038